5 Troubleshooting
1006-0836-000 02/03
5-11
Paw < -10 cm H
2
O
User Alarm
Subatmospheric pressure
(<-10 cm H
2
O).
Check patient condition,
spontaneous activity?
Increase fresh gas flow. Look
for high flow through gas
scavenging.
Calibrate the flow sensors. With
active scavenging, check the
negative relief valve on the
receiver.
PEEP Valve (DAC)
Failure
Error
Minimum
Monitoring
Indicates a failure of the
control circuit for the
PEEP Valve.
Ventilate manually.
Replace:
• Control Board.
PEEP Valve (Voltage)
Failure
Error
Minimum
Monitoring
Indicates a problem with the
PEEP Valve or the
connections to the PEEP
Valve.
Ventilate manually.
Check/Replace:
• PEEP Valve
• PEEP Valve
harness
• PEB/VEB
• Cable to CB/J14
• Control Board.
PEEP Safety Valve
(Drive) Failure
Error
Minimum
Monitoring
Indicates a higher than
allowed current draw on
the PEEP Safety Valve.
Ventilate manually.
Check/Replace:
• PEEP Safety Valve
• PEEP Safety Valve
harness
• PEB/VEB
• Cable to CB/J14
• Control Board.
PEEP Safety Valve
Failure
Error
Minimum
Monitoring
Indicates that the powerup
test of the PEEP Safety
Valve failed.
Ventilate manually.
Check/Replace:
• PEEP Safety Valve
• PEEP Safety Valve
harness
• PEB/VEB
• Cable to CB/J14
• Control Board.
Pinsp Not Achieved
User Alarm
Indicates a problem with
breathing circuit
connections or that the
ventilator is unable to
deliver requested
pressure to the patient.
Check breathing circuit
connections. Check settings.
Pressure Monitoring
Channel Failure
Error
Minimum
Monitoring
Indicates a problem with
patient airway
overpressure monitor.
Ventilate manually.
Check/Replace:
• PEB/VEB
• Cable to CB/J14
• Control Board.
Message
Alarm/Error
Cause
User Action/Concerns
Service Repair