7100 Anesthesia Ventilator
4-10
02/03 1006-0836-000
Heliox Mode
Enabled
— Disabled
Only Aestiva machines can be configured to deliver Heliox. Aestiva machines
that include the Heliox option, should have the Heliox Mode “Enabled”. With
the Heliox Mode enabled, the operator can choose to turn the Heliox Mode On
or Off (on the Setup menu).
All Aespire machines should have the Heliox Mode disabled.
If the machine is not configured to deliver Heliox, the Heliox Mode should be set
to “Disabled”. With the Heliox Mode disabled, the user’s “Setup” menu will not
include the Heliox option.
Whenever you enable or disable the Heliox Mode, the screen will display the
following warning:
Note
A similarly worded warning will appear when you change the “O
2
Monitoring,”
“Volume Monitoring,” and the “Volume Compensation” settings.
√
E
Alarm Limits
Automatic
— User Adjustable
System Type
Aestiva
Currently, the 7100 ventilator is available in the Aestiva machine and the
Aespire machine. Since the 7100 ventilator functions identically in either
machine, the System Type setting is displayed as “Aestiva.” The cursor will skip
over this selection.
Flow Sensor Correction
Lead Filter Off
— Lead Filter On
This setting relates to the MIA/VMB board. For current MIA/VMB assemblies,
this setting should be
Lead Filter Off
.
O
2
Monitoring
Enabled
— Disabled (if site preference)
May not appear in some configurations.
Volume Monitoring
Enabled
— Disabled (if site preference)
May not appear in some configurations.
If “Volume Monitoring” is set to “Disabled,”
“Volume Compensation” is automatically set to “Disabled.”
Volume Compensation
Enabled
— Disabled (if site preference)
May not appear in some configurations.
Pressure Mode
Enabled
— Disabled (if site preference)
May not appear in some configurations.
Pressure Waveform
Enabled
— Disabled (if site preference)
May not appear in some configurations.