7 Repair Procedures
1006-0836-000 02/03
7-21
To replace
the Serial Adapter Board
1. Disconnect the two cables from the bottom of the SAB:
•
Cable (
A
) from machine switches
(system switch, O
2
flush, O
2
supply).
•
Cable (
B
) from control module.
2. Disconnect the serial output cable (
C
).
3. Remove the three screws (
D
) that hold the SAB to the back wall.
4. Remove the three nuts (
E
) that hold the SAB mounting bracket to the
enclosure.
5. Reassemble in reverse order.
A
B
(SAB)
C
E
D