![Datascope Passport 2 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/datascope/passport-2/passport-2_service-manual_3701710040.webp)
Patient Connector Board AAMI
Theory of Operation
2 - 30
0070-10-0441
Passport 2®/Passport 2 LT™ Service Manual
2.5.7
0670-00-0682-01
This module is responsible for acquisition of most of the patient-safety-isolated signals,
namely the ECG, Temperature and two channels of Invasive Blood Pressure connectors IBP1
and IBP2. The Connector board collects all of the individual patient cables into a single
ribbon cable, then passes the signals to the main board.
2.5.8
0670-00-0682-02
This module is responsible for acquisition of the patient-safety signals, namely the ECG and
the Temperature. This module is without the dual 6 pin connectors for the IBP. The Connector
board collects all of the individual patient cables into a single ribbon cable, then passes the
signals to the main board.