![Datascope Passport 2 Скачать руководство пользователя страница 144](http://html1.mh-extra.com/html/datascope/passport-2/passport-2_service-manual_3701710144.webp)
Preventive Maintenance Schedule
Preventive Maintenance
7 - 2
0070-10-0441
Passport 2®/Passport 2 LT™ Service Manual
7.1
Preventive Maintenance Schedule
The following is a list of activities required for periodic maintenance of the
Passport 2
monitor. The physical inspection, replacement of consumable items and performance checks
should be performed at the recommended intervals stated below. Mindray DS USA, Inc. is
not responsible for component failure or loss resulting from the use of stated consumable
items beyond their recommended replacement interval.
7.2
Mechanical / Physical / Visual Inspection
7.2.1
Perform at Twelve Month Intervals
Suggested Inspections for Wear and Abuse:
1.
Outer case, Line Cords, Rolling Stands, Wall Mounts, Modular Accessories and
Interconnecting Cables.
2.
Patient Interface Connections (ECG, IBP, SpO
2
, Temp, CO
2
and NIBP).
7.3
Preventive Maintenance Kit
1.
N/A
7.4
Perform Verification and NIBP Calibration –
Annually
1.
See section 6.5., “Verification”.
7.5
Perform Verification and CO
2
Calibration – Annually
1.
See Test Outline in Service Manual.
2.
In order to provide accuracy verification of the Microstream CO
2
, calibration is required
every 4000 operating hours or once a year, whichever comes first. In addition, to
achieve the most accurate CO
2
readings possible, it is advised that the first calibration
be performed after 1200 operating hours.
3.
Replace Pump and Scrubber after 20,000 Operating Hours or as Required.
4.
Replace the CO
2
Sensor Assembly after 20,000 Operating Hours or as required.
7.6
User Preventive Maintenance Introduction
This section of the manual outlines routine maintenance that should be performed by the user.
The
Passport 2
Monitor is designed for stable operation over long periods of time and
under normal circumstances should not require technical maintenance
beyond that described in this section. However, it is recommended that routine maintenance
calibration and safety checks be performed at least once a year, or more often as required
by local statutory or hospital administration practice.