Diagnostics
Calibration Procedure
6 - 14
0070-10-0441
Passport 2®/Passport 2 LT™ Service Manual
6.4.7.5
Dump Test (Adult, Neonate)
The purpose of this test is to verify the valve that control the and dump rate is functioning
properly.
1.
Connect the dummy cuff / test chamber to the side panel fitting.
2.
Rotate the control knob to highlight the dump test. Once the dump test is highlighted
press the control knob to activate the test.
3.
The chamber will inflate to 270mmHg of pressure (Adult) 170mmHg (Neonate). The
dump valve will start to deflate at 260 (Adult) 150 (Neonate) after 10 seconds (Adult) 5
seconds (Neonate) the unit will dump the pressure to approximately 15mmHg (Adult)
5mmHg (Neonate). On screen the result of the test will be viewed.
FIGURE 6-16
Specification: Dump Rate – 260 to 15mmHg / 10sec or less – Adult – 500cc
test chamber
Dump Rate – 150 to 5mmHg / 5sec or less - Neonate – 500cc test chamber
Specification: Dump Rate – 260 to 15mmHg / 14sec or less – Adult – 700cc
test chamber
Dump Rate – 150 to 5mmHg / 7sec or less - Neonate – 700cc test chamber