Verification
Calibration Procedure
6 - 18
0070-10-0441
Passport 2®/Passport 2 LT™ Service Manual
b.
NIBP
Set Start Pressure - 180 mmHg
Interval - 5 minutes
IBP1 - Scale 0 to 160mmHg
IBP2 - Scale 0 to 80mmHg
c.
SpO
2
Averaging mode - 2
Sensor Off Audio - off
d.
CO
2
- (optional)
Apnea Delay - 30
Scale 40 mmHg
e.
Respiration -
Resp lead - II
Apnea Delay - 30
Resp source - Auto
Scale - 3
f.
Gases - (optional)
Select agent - Auto
O
2
scale - 100%
N
2
0 Scale - 10%
6.5.2
ECG Tests
6.5.2.1
Initialization
1.
Observe that the trace display sweeps across the waveform 1 screen in five seconds.
There should be five complete ECG cycles. The same display and timing should be seen
in the Waveform 2 screen.
2.
Check the following sweep speeds for the appropriate displays:
12.5 mm/sec – 10 second sweep/window.
6.5.2.2
Leads OFF
1.
Disconnect one lead at a time RA, RL, LL, LA, and C from the simulator and observe that
the message “Lead OFF” appears on the display.
2.
Set the ECG simulator to Short Leads. Verify the resolution does not exceed one pixel
resolution.