Diagnostics
Calibration Procedure
6 - 12
0070-10-0441
Passport 2®/Passport 2 LT™ Service Manual
6.4.7.3
Motor Pump Test
1.
Connect the dummy cuff / test chamber to the side panel fitting.
2.
Rotate the control knob to highlight the motor pump test. Once the motor pump test is
highlighted press the control knob to activate the test.
3.
On screen the target pressure of 300mmHg will be view on screen. The time required to
pump to 300mmHg will be viewed on screen.
FIGURE 6-14
Specification: Pump to 300mmHg in < 35.0 seconds. – 500cc test chamber
Specification: Pump to 300mmHg in < 49.0 seconds – 700cc test chamber