Passport 2®/Passport 2 LT™ Service Manual
0070-10-0441
6 - 9
Calibration Procedure
Diagnostics
i.
Select down to NIBP Overpressure Voltage. Adjust potentiometer R 31 on the NIBP
Module and verify on screen the Overpressure Voltage is 0.100 volts.
FIGURE 6-11
j.
Reassemble unit and verify proper operation.
B.
Specification with P/N 0670-00-0730/0670-00-0746-01
No manual adjustment is required. The voltage setpoint is controlled by software. If “Check
Calibration/Cuff Overpressure” appears on the screen in message area, replace the NIBP
module accordingly.
C.
Specification: .030 to .170 volts