6.4.7.2
Static Pressure Calibration
The purpose of this test is to verify the pressure transducer sensitivity for optimal accuracy.
FIGURE 6-12
1.
Connect the dummy cuff / test chamber to the side panel fitting.
2.
Rotate the control knob to highlight the static pressure calibration. Once the static
pressure calibration test is highlighted press the control knob to activate the test.
3.
Using a bulb and valve manually inflate the chamber to 150mmHg. (150mmHg is the
middle of the specified range)
4.
Verify the pressure displayed on screen matches the pressure viewed on a Digital
Mercury Manometer.
Specification: 0 to 3/- 3mmHg
If the readings do not match follow the following procedure.
1.
Turn unit off.
2.
Remove the eight screws from the rear of the unit and separate the front and rear
housing.
3.
Remove the NIBP Pump assembly and bracket from the Main Assembly. Be sure to
reattach NIBP pump cable to module before next step.
4.
Locate R81 on the NIBP Module. Turn unit on and enter the Diagnostics as stated in
section 6.4. Reenter the diagnostics menu and select the Static Pressure Calibration
menu.
5.
Manually inflate the chamber to 150mmHg.