Passport 2®/Passport 2 LT™ Service Manual
0070-10-0441
6 - 21
Calibration Procedure
Verification
6.5.6
NIBP Verification
1.
Connect the Adult cuff connector to the NIBP hose. Attach the NIBP hose to the Cuff
connection the left side on the monitor.
2.
Apply cuff and press the start key on the
Passport 2
.
3.
Verify the pump motor starts to pump and inflates the cuff to 180 mmhg (Adult). The cuff
will begin to deflate and obtain a blood pressure reading of Sys/ Dia/ Mean in about
20 to 30 seconds after peak pressure is obtained.
4.
Verify the reading on screen.
6.5.7
Battery Operation Verification
1.
If batteries are installed in the unit, remove them.
2.
Verify the unit functions on Line power correctly.
3.
Install the two batteries in the appropriate slots located on the left side of the monitor.
4.
Remove the line cord from the unit. Verify the unit operation is not interrupted.
5.
Remove one of the batteries and verify the unit still operates. Verify the second battery
operates in the same manner by reinstalling the first battery and removing the second
battery.
6.5.8
CO
2
Operation Verification
1.
Connect the Filterline Short Term assembly to the input port of the CO
2
connector on the
left side of the monitor.
2.
Attach a can of Calibration Gas P/N 0075-00-0033-01 to the Filterline Short Term
assembly. Feed gas into monitor and verify the ETCO
2
, Inspired CO
2
and respiration
readings occur on the screen.
6.5.9
Leakage Current Tests
1.
Plug the line cord of the unit into the safety analyzer. Connect the case ground lead of
the analyzer to the equipotential lug of the monitor on the rear of the monitor.
2.
Perform the tests under the following conditions:
a.
Case Grounded:
Normal polarity
Normal polarity with open neutral
b.
Case ungrounded:
Normal polarity
Normal polarity with open neutral
Reverse polarity
Specs: Verify the current reading of the test is less than 100 µA under normal operating
conditions.
Less than 300 µA under a single fault condition for 120 VAC and less than 500 µA under a
single fault condition for 230 VAC.