Maintenance ■ Pick and Place Probe
- 104 -
Data I/O ■ 096-0465-001C
Note:
Do not touch the probe tip with bare hands. Finger oils can cause
small devices to stick.
Adjusting Probe Blow-Off Pressure
Note:
The flow controls are set at the factory and should not require
adjustment. If adjustments need to be made, they should be done in small
steps until the desired results are reached.
Blow-off
is a small puff of air applied at the probe during the device drop
event to assist in removing a device from the probe.
Blow-off
is produced
by a vacuum generator on the PNP head. If set too high,
blow-off
could
cause device misalignment during placement.
T
OOLS
R
EQUIRED
• Large flat screwdriver
• Very small flat screwdriver
To adjust blow-off pressure on any vacuum generator:
1.
Click System > Misc. I/O.
2.
At the I/O Interface window, select the desired vacuum generator,
probe 1 or probe 2 (or both).
3.
Click Vacuum to the OFF position. Refer to Figure 4-4.
4.
Click Blow-off to the ON position.
Figure 60:
Vacuum and Blow-off on the I/O Interface window.
WARNING: Possible collision hazard! The high speed and force of
the gantry can seriously harm anyone working inside the
Содержание PSV5000
Страница 12: ......
Страница 36: ......
Страница 160: ......
Страница 161: ...Electrical Schematics for PSV5000 and Equipment ...
Страница 162: ......
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 176: ...PROGRAMMABLE MEDIA EXPERTS ...