Running a Job on PSV5000 ■ Removing Devices
- 64 -
Data I/O ■ 096-0466-001E
Removing Devices
Remove unwanted devices in programmer sockets by:
Manually
1.
Pause or
Finish
(end) a job.
2.
Right-click on the desired
programmer to open a
submenu, and then click
Opener UP.
3.
Using ESD precautions,
open a safety door and re-
move the devices with a
vacuum tweezer.
At the Run Window
1.
Pause
or
Finish
(end) a job.
2.
Click Clear. Device will be
moved to the reject media.
Stopping the System
There are three methods for stopping the machine:
•
Emergency Stop
—
Pushing any of the red E-Stop buttons stops the
movement of the PNP head immediately in an emergency. (Power to the
gantry motors is disconnected and communication to the Laser head is
disconnected.) The CH700 Application must be closed and the job re-
started in TaskLink.
•
Pausing a Job
—
Clicking
Pause
in the Run window
—
stops the PNP head
at its next destination. Devices finish programming if started but are not
removed from sockets. The programming session can be continued.
•
Finish a Job
—
Clicking
Finish
in the Run window
—
stops picking blank
devices, finishes the current programming cycle, and removes devices
from all the sockets placing them into appropriate media. The pro-
gramming job can be resumed.
Job throughput
as displayed in the Run window is af-
fected by the length of time a job is paused.
Nominal throughput