Running a Job on PSV5000 ■ List of Steps to Start a Job
- 48 -
Data I/O ■ 096-0466-001E
Teaching the Tray 1 Location
1.
In the
CH700 Setup
window, click
System
(in the right pane).
2.
In the
System
window, click
Gantry
.
Figure 50:
Click ‘Gantry’ for handler functions.
3.
In the
Gantry
window, click
Tr1
to move PnP head to the Tray 1 location.
Figure 51:
Click ‘TR1’ to move PnP head to Tray 1 location.
4.
Center Probe 1 over the device (use X and Y adjustment arrows), and
then click
Z AutoFind
.
Figure 52:
Click ‘Z AutoFind’ to teach the vertical height.