AP Blot - SERVICE MANUAL
SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 63
OF
Turns off all motors
P2A
Turns on waste pump
P2S
Turns off waste pump
PA
Turns on aspiration pump, waste pump 2 and opens aspiration electro-valve on the wash basin
PB1
Creates a logic signal of 100 msecs. on busy 1.
PS
Turns off peristaltic pump, waste pump and turns off waste electro-valve, connecting it with the
aspiration probe.
PVn
Sets peristaltic pump rotation speed.
RAc,v
Sets the semi-period ‘v’ in microseconds in ‘c’ variable of ramp table.
RB1
Puts logic level 0 the busy signal 1.
RCc,v
Sets the semiperiod ‘v’ in microseconds in ‘c’ variable of ramp table..
RMc n
repeats ‘c’ macro for ‘n’ times (0
≤
n
≤
65535)
RN
Displays command echo and serial number (format = 10 characters)
RSa,b,c
Defines brake ramp where ‘a’ = start frequency, ‘b’ = stop frequency and ‘c’ = number of points along
ramp. Valid for both syringes
RXa,b,c
Calculates acceleration and brake ramp, where ‘a’ = start frequency, ‘b’ = stop frequency and ‘c’ =
number of ramp elements (default a = 125, b = 2500, c = 240).
RYa,b,c
Calculates acceleration and brake ramp, where ‘a’ = start frequency, ‘b’ = stop frequency and ‘c’ =
number of ramp elements (default a = 400, b = 4000, c = 250).
RZa,b,c
Defines acceleration ramp and brake, where ‘a’ = start frequency, ‘b’ = stop frequency and ‘c’ = number
of ramp elements valid for Z1 and Z2 axis (default a = 200, b = 4000, c = 200).
SA0
Disables bottle alarm
SA1
Activates display warning whenever any bottle changes status.
SB1
Puts logic level 0 the busy signal 1.
SE0
Disables instrument pause if a liquid error is detected (break on error).
SE1
Activates instrument pause and returns axis Z1 to zero position if break on error is detected.
SE2
Disables break on error and liquid error message
SGA{a..z}n
Sets aspiration speed for both probes (default 3000hz).
SGB{a..z}n
Sets bubble volume valid for both Z axis.
SGC{a..z}n
Sets combined speed for both Z axis.
SGD{a..z}n
Sets dispensation speed for both probes (default 3000hz).
SGE{a..z}n
Sets liquid alarm status (see SE0, SE1, SE2 commands).
SGZ1{a..z}n
Sets Z1 maximum height
SGZ2{a..z}n
Sets Z2 maximum height
ST
Takes timer 1 to zero and restarts
SUL
Corrisponds to how long valve is open (in milliseconds) for probe to dispense 1ml
SVAn1
Sets probe speed in aspiration phase, ‘n1’ sets probe 1 motor speed (default 3000 Hz).
SVDn1
Dispensation phase probe speed, ‘n1’ sets probe 1 motor speed (default 3000 Hz).