3.3.1.8.
Dilutor unit hydraulic component diagram
The table below describes the different functions of the dilutor unit hydraulic components.
The diagram shows how the dilutor unit circuit is set up to work.
WASH bottle
Contains probe-washing liquid, 5 litre capacity
Level Sensor
Sensor for the three different bottles and connecting cables and connectors
Probe washing pump
Peristaltic pump A for washing probe, internally and externally, driven by step motor
Probe wash basin
For external probe wash and draining any excess liquid
Syringe
For aspirating and dispensing liquid through probe
Peristaltic Pump A
Peristaltic Pump B
Syringe
Dispensation probe
Aspiration probe
Needle washing basin
W ASTE bottle
W ASH BUFFER
bottle
WASH bottle
W ASH
W ASH BUFFER
W ASTE
3 way
Electrovalve
3 way Electrovalve