AP Blot - SERVICE MANUAL
SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 46
3.5.4. SUBSTITUTING Z
1
and Z
2
AXIS PARTS
Substituting Z1 Axis probe
Para. 3.5.4.1
Substituting Z1 Axis probe motor
Para. 3.5.4.2
Substituting Z2 Axis probe
Para. 3.5.4.3
Substituting Z2 Axis probe motor
Para. 3.5.4.4
3.5.4.1.
Substituting Z
1
Axis probe
a) Remove instrument cover
b) Remove protective carter by unscrewing two fixing screws
c) Detach AWG15 tubing from Probe
d) Remove liquid sensor cable connector from INTEROP2 PCB
e) Remove screw that fixes probe to stand
f)
Replace probe with new probe and tighten fixing screw
g) Reconnect probe and liquid sensor connector (NON CE SU VERSIONE ITaliana) to INTEROP2 PCB and
reposition AWG15 tubing to Probe
h) Check Z axis settings: Z steps and LT parameter
i)
Check X-Y centering
2 fixing
screws
fixing screw