AP Blot - SERVICE MANUAL
SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 31
3.3.3. SUBSTITUTING DILUTOR UNIT COMPONENTS
To substitute syringe and syringe plunger refer to para. 4.8.5.
The following table lists procedures and paragraphs relevant to these substitutions.
Substituting valve – syringe – dispensation probe tubing
Ref. para. 3.3.3.1
Substituting valve mechanism
Ref. para. 3.3.3.2
Substituting 3-way Electro valve
Ref. para. 3.3.3.3
Substituting 3-way valve mechanism
Ref. para. 3.3.3.4
Substituting Peristaltic Pump Tubing
Ref. para. 3.3.3.5
Substituting Dilutor Motor
Ref. para. 3.3.3.6
Substituting Opto SIDILOP PCB
Ref. para. 3.3.3.7
Substituting INTERFA2 PCB
Ref. para. 3.3.3.8
Substituting probe wash basin
Ref. para. 3.3.3.9
3.3.3.1.
Substituting valve – syringe – dispensation probe tubing
a) Disconnect AWG15 tubing from probe
b) The old tubing (to be replaced) should be used as a guide or ‘driver’/track
c) Join the ends of the old and new tubing together with adhesive tape.
d) Slowly pull old tubing.
e) Separate new tubing from the original “driver” tubing and join the new tubing to the valve unit joint and probe.
AWG15
tubing