AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 80
4.7.4. Movement Test
From “
Main Functions
” pull down menu select “
Movement Test
” to check that the
different moving parts are functioning correctly
This opens the following display:
Click on “
Start
” to run test.
The test automatically checks movement in the X-Y-Z axis, the shaking plane movement, primes hydraulic circuit and
checks communication with the camera.
If all the controls are positive at the end of the test, the following display appears.