AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 102
4.8.9. Calibrate camera
From “
Dilutor
” pull down menu select “
Calibrate camera
” .
The following display will appear:
Fig. 1
Test correct gain calibration
1) Check that internal light is on, instrument cover is closed and top cover is inserted.
2) Select “
Auto Gain ON
”. The camera will then move to frame the Calibration Strip (as shown in Fig 1 ) and
will automatically set a white balance.
3) After this balance check Gain Value.
4) If Gain Value is between 40 and 60 - the diaphragm is perfectly calibrated. If not then Calibrate Gain.
5) Select “
Auto Gain OFF
” to interrupt test