AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 86
Calibration
X-axis Calibration
Centering the Dispensation Probe along X-axis over the Work Area is a mechanical calibration procedure where the
OPTO PCB with the optic sensors is moved together with the
APX-XGOLAV-X
on the X carriage at the back of the
device as shown in fig. 1.
1. Remove device top cover
2. Loosen fixing screws on APX-XGOLAV-X PCB
3. Move PCB manually to bring the probe directly over the center of the reference point and then tighten the
screws on the PCB
4. Select “
Go up
” to lift probe up and then click on icon to turn off motors
5. Move arm that controls optic sensors so that it now moves freely over the encoder
6. Return to Home position
7. Select “
Go down
” and check that probe is correctly in position
8. Repeat this procedure from point (2) until position is correctly centered
XGOLAV-X
PCB
Fixing screws