AP Blot - SERVICE MANUAL
SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 33
3.3.3.5.
Substituting Peristaltic Pump Tubing
a) Turn on instrument. Run XCalib program. From “Dilutor” menu select “Prime”
b) Detach wash loading and buffer tubing from their bottles
c) Set prime for at least 60 seconds to empty hydraulic circuit. Select ‘Prime’
d) After this prime rinse, detach tubing and then slip tubing off pumps.
B
A
e) Place new peristaltic tubing into position and re-connect it up following the diagram sticker on device.
f)
Reconnect both Wash loading and Buffer tubing to their respective bottles.
g) Run another prime by selecting “Prime” option to rinse hydraulic circuits again
White
/ Black
Blue /
White
Peristaltic pump tubing