AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 94
Check horizontal position:
•
Click on “
Move up
” and then “
Reset
” to return shaking plane to the Home (flat) position
•
Use a spirit level to check that shaking plane is level horizontally along both X- and Y-Axis
If uneven or not level, proceed with Calibration procedure described here:
Calibration
Remove all strip holder plates from shaking plane and use a spirit level to check that shaking plane is level, both for X
and Y axis
Regulate Swing PCB opto (under shaking plane) if not level (ref. para.3.3.6.1.).
Horizontal Position Calibration
•
Remove Plate Holder (Fig.X)
•
Life off shaking plane and loosen fixing screws of the OP-Swing PCB (Fig. X)
•
Move the PCB to mechanically regulate the shaking plane home position
•
Lower the shaking plane and repeat the procedure to check horizontal level of shaking plane
•
When this procedure is completed replace the strip holder plate correctly
Strip holder
incubation
tray
OP-Swing PCB