Cooling Water Inlet Connection
The cooling water inlet connection provides water to the compressor package from your
facility to cool the compressor package during operation. The connector thread size is 1/4
FPT (1/4” Female National Pipe Thread).
The water must meet the specifications provided in the Cooling Water Specifications
table in Section 5.2.
Cooling Water Outlet Connection
The cooling water outlet connection carries heated water away from the compressor
package after the water has been heated by cooling the compressor package during
operation. The connector thread size is 1/4 FPT (1/4” Female National Pipe Thread).
Pressure Switch Reset (Low Pressure)
The compressor package has a low-pressure shutdown switch that is set at 35 ± 5 PSIG
(2.4 ± .34 bar). The pressure switch manual reset button is located on the bottom of the
front panel as shown in Figure 5-3 or 5-4.
Temperature Switch Reset (High Temperature)
The high temperature shutdown switch is set at 120°F (49°C). The high temperature
manual reset button is located on the bottom of the front panel as shown in Figure 5-3 or
5-4. The temperature switch cannot be reset until the sensor has cooled by
approximately 15°F (8.5°C).
Power Cord
The power cord supplies power from the wall to the entire system.
Содержание LNP-40
Страница 1: ...Model LNP 40 Liquid Nitrogen Plant INSTALLATION OPERATION and ROUTINE MAINTENANCE MANUAL...
Страница 4: ...1 1 N Section 1 Overview...
Страница 10: ...2 1 N Section 2 Warranty...
Страница 13: ...3 2 N Section 3 Safety...
Страница 25: ...4 1 N Section 4 Inspection and Unpacking...
Страница 29: ...5 3 N Section 5 Specifications...
Страница 50: ...6 1 N Section 6 Installation...
Страница 67: ...7 2 N Section 7 Operation...
Страница 87: ...8 1 N Section 8 Routine Maintenance...
Страница 98: ...Appendix Spray Lubricant Instructions...