Safety
3-9
Section 7.7.3 Dewar
CAUTION
A dewar pressure greater than 25 PSIG will destroy the level indicator/switch.
3.2.4 Section
8.
Routine maintenance
Section 8.1 Introduction
CAUTION
Failure to maintain the system properly will void the warranty.
Section 8.3 Cold head
CAUTION
Service to the cold head must be performed using Cryomech parts and the
appropriate work instructions.
CONTACT CRYOMECH IF THE COLD HEAD NEEDS SERVICING.
Section 8.4 Helium compressor adsorber replacement
CAUTION
At no time should the Aeroquip
®
couplings be removed from the adsorber when
replacing the adsorber. Replacement can be completed without relieving system
pressure since the adsorber is equipped with Aeroquip
®
couplings for sealed
removal.
WARNING
The input power must be disconnected from the helium compressor package before
removing any helium compressor package panels.
Section 8.5 Vent excess helium
CAUTION
Venting more than 5 PSIG (.34 bar) of helium per minute will lead to improper oil
migration within the system. If this condition occurs, factory service will be required.
10Jun11
Содержание LNP-40
Страница 1: ...Model LNP 40 Liquid Nitrogen Plant INSTALLATION OPERATION and ROUTINE MAINTENANCE MANUAL...
Страница 4: ...1 1 N Section 1 Overview...
Страница 10: ...2 1 N Section 2 Warranty...
Страница 13: ...3 2 N Section 3 Safety...
Страница 25: ...4 1 N Section 4 Inspection and Unpacking...
Страница 29: ...5 3 N Section 5 Specifications...
Страница 50: ...6 1 N Section 6 Installation...
Страница 67: ...7 2 N Section 7 Operation...
Страница 87: ...8 1 N Section 8 Routine Maintenance...
Страница 98: ...Appendix Spray Lubricant Instructions...