Specifications
5-3
5.2.3 Helium compressor electrical specifications
Parameter
200/230 Volt
60 Hz Model
440/480 Volt
60 Hz Model
200 Volt
50 Hz Model
380/415 Volt
50 Hz Model
Nominal
voltage
200/230 VAC
440/480 VAC
200 VAC
380/415 VAC
Operating
voltage
range
180 - 253 VAC
396 - 528 VAC
180 - 220 VAC
342 - 456 VAC
Frequency
60 Hz
60 Hz
50 Hz
50 Hz
Phase 3
3
3
3
Water
Cooled
Input Power
Maximum:
Steady
state:
6.4 kW
5.3 kW
N/A kW
N/A kW
N/A kW
N/A kW
5.2 kW
4.4 kW
Current
21 A
N/A A
N/A A
9 A
Air Cooled
Input Power
Maximum:
Steady
state:
6.4 kW
5.9 kW
N/A kW
N/A kW
5.0 kW
4.3 kW
5.6 kW
4.7 kW
Current
21 A
N/A A
16.5 A
10 A
Dedicated
circuit
breaker
40 A
25 A
40 A
25 A
Mains
supply
voltage
fluctuations
Up to ± 10% of
the nominal
voltage
Up to ± 10% of
the nominal
voltage
Up to ± 10% of
the nominal
voltage
Up to ± 10% of
the nominal
voltage
5.2.4 Helium compressor fuse specification
System
Voltage
Transformer
Fuse*
Fuse Rating
200/230 VAC
200 VAC
250 VA
Primary
2.5 A
440/480 VAC
380/415 VAC
250 VA
Primary
1.25 A
*Fuse is 13/32” x 1-1/2”, Type CC time delay fuses, rated for 600V.
25Oct11
Содержание LNP-40
Страница 1: ...Model LNP 40 Liquid Nitrogen Plant INSTALLATION OPERATION and ROUTINE MAINTENANCE MANUAL...
Страница 4: ...1 1 N Section 1 Overview...
Страница 10: ...2 1 N Section 2 Warranty...
Страница 13: ...3 2 N Section 3 Safety...
Страница 25: ...4 1 N Section 4 Inspection and Unpacking...
Страница 29: ...5 3 N Section 5 Specifications...
Страница 50: ...6 1 N Section 6 Installation...
Страница 67: ...7 2 N Section 7 Operation...
Страница 87: ...8 1 N Section 8 Routine Maintenance...
Страница 98: ...Appendix Spray Lubricant Instructions...