Installation
6-4
6.3.3 Connect the helium compressor package to the main power
CAUTION
A voltage deviation of more than 10% above or below the voltage rating can cause
compressor motor overheating and possible failure.
Indications of operation outside
that voltage range will void the compressor warranty
.
1) The system MUST be connected to a dedicated circuit breaker. The breaker must be
mounted near the helium compressor package, within easy reach of the operator,
and must be marked as the disconnecting device for the system. Specifications for
circuit breakers vary according to the system’s operating voltage. See the electrical
specification tables in Section 5 for more information.
2) The helium compressor package comes with a main power cord attached. Assure
that the length of the cord is sufficient to safely connect to the power source. If the
cord is not sufficiently long, adjust the location of the helium compressor package.
WARNING
Be sure to follow all local electrical codes and guidelines.
3) Make sure that the dedicated circuit breaker is turned OFF.
4) The ground (or earth) wire in the power cord is either green (60 Hz systems) or
green/yellow stripe (50 Hz systems). Connect the ground wire in the power cord to
the ground (or earth) connector in the breaker panel, making sure to tighten the wire
into the connector securely. It is important not to disable this wire.
5) Connect the remaining hot wires in the power cord to the corresponding lugs on the
dedicated breaker in the breaker panel, making sure to tighten the connector
securely. The order of the wires is not important for single phase, CP800 Series
compressor packages. For 3 phase compressor packages, the order of the wires is
not important at this time - correct order will be determined in Section 7.
WARNING
One lead of the helium compressor package is grounded. Never bypass this ground
or attach the helium compressor package to an ungrounded circuit. A dangerous
electrical hazard will develop.
6.3.4 Connect the helium flex lines to the cold head and the helium
compressor package
CAUTION
Follow the procedure carefully when connecting and disconnecting the helium flex
lines. Failure to follow the procedure can cause accidental coupling disassembly,
destruction of the sealing O-ring, and helium loss.
Содержание LNP-40
Страница 1: ...Model LNP 40 Liquid Nitrogen Plant INSTALLATION OPERATION and ROUTINE MAINTENANCE MANUAL...
Страница 4: ...1 1 N Section 1 Overview...
Страница 10: ...2 1 N Section 2 Warranty...
Страница 13: ...3 2 N Section 3 Safety...
Страница 25: ...4 1 N Section 4 Inspection and Unpacking...
Страница 29: ...5 3 N Section 5 Specifications...
Страница 50: ...6 1 N Section 6 Installation...
Страница 67: ...7 2 N Section 7 Operation...
Страница 87: ...8 1 N Section 8 Routine Maintenance...
Страница 98: ...Appendix Spray Lubricant Instructions...