Warranty
2-2
•
The air compressor must be installed and operated according to the manufacturer’s
specifications included in the air compressor’s instruction manual.
Failure to follow
the manufacturer’s instructions will void the warranty.
See Sections 6 and 7.
•
If the air compressor must be bolted to the floor, please pay special attention to the
mounting instructions included in Section 3 of the air compressor’s instruction
manual.
Failure to mount the air compressor according to the manufacturer’s
instructions will void the warranty.
See Sections 6 and 7.
•
Do not allow the nitrogen flex line to come into contact with corrosives or any type of
commercial cleaning agent.
Leaks caused by exposure to corrosives or commercial
cleaning agents will not be covered under warranty.
See Section 6.
•
Do not bend the nitrogen flex line to less than 5 inch (12 cm) radius or permanent
damage may occur.
This type of damage is not covered under the warranty.
See
Section 6.
•
CP900 SERIES COMPRESSORS ONLY.
Running the compressor package in
reverse for any significant length of time will void the warranty.
If the power cord is
not wired properly the compressor will run in reverse and there will be no pressure
differential on the pressure gauges. Turn the system off immediately. Follow the
instructions in Step 5, below, to rewire the power cord correctly. See Section 7.
•
If excessive amounts of condensate are allowed to accumulate in the filter bowls, the
condensate will move downstream and destroy the N
2
generator.
This type of
damage is not covered under the warranty.
See Section 7.
•
Failure to maintain the system properly will void the warranty.
See Section 8.
•
When adding helium, the helium must be 99.999% pure. Contamination by other
gases will result in the freezing of the contaminant gases in the cold head because
their freezing temperature is much higher than that of helium. Contaminants in the
helium charge will severely degrade the cold head’s function and it will require factory
servicing.
Contamination of the helium by other gases is a common cause of
premature failure and, unless resulting from a system failure, is not covered by the
warranty.
See Section 8.
•
Never wet either part of the system.
Water getting into the system will void the
warranty.
See Section 8.
14Feb07
Содержание LNP-40
Страница 1: ...Model LNP 40 Liquid Nitrogen Plant INSTALLATION OPERATION and ROUTINE MAINTENANCE MANUAL...
Страница 4: ...1 1 N Section 1 Overview...
Страница 10: ...2 1 N Section 2 Warranty...
Страница 13: ...3 2 N Section 3 Safety...
Страница 25: ...4 1 N Section 4 Inspection and Unpacking...
Страница 29: ...5 3 N Section 5 Specifications...
Страница 50: ...6 1 N Section 6 Installation...
Страница 67: ...7 2 N Section 7 Operation...
Страница 87: ...8 1 N Section 8 Routine Maintenance...
Страница 98: ...Appendix Spray Lubricant Instructions...