F
IBER
S
CAN
ONE
™
Select the “
CAL
” option form the setup menu, on the instrument the following
screen is shown:
Figure 22 - Calibration main screen
The instrument verifies the calibration by pressing the measurement key. At
the end of the procedure, the following screen will be shown, with two
possibilities:
•
If the measured value fits between the 5% reference tolerance, the
firmware routine automatically will exit the calibration menu, after
approx. 5 seconds.
Figure 23 - Instrument is already calibrated
•
If the measured value is out of the 5% reference tolerance (positive or
negative), a message will request to perform the calibration process.
FiberScanONE™
User
manual
rev.
1.0.3
Page
38 of 74