F
IBER
S
CAN
ONE
™
6.5.3 Firmware update
FiberScanONE™ internal firmware can be easily upgraded to the latest version,
to add i.e. new functions and improve the instrument flexibility.
To enable firmware update, the ENTER key must be kept pressed down for at
least 5 sec.
Once the BSL connection is enabled, the following image will appear on the
screen.
Figure 35 - Firmware upgrade screen
Once the connection is established, the control unit's firmware can be updated
by connecting it to a PC equipped with the appropriate update software.
For detailed instruction please refer to the software update manual.
WARNING!
This operation must be carried out only by expert personnel!
Always consult our Direct Service at +39 035 715282. Wrong
procedures might damage the control unit ROM
WARNING!
Once the BSL function has been activated, the control unit will no
longer accept any commands and cannot be started up again except
by the software update
WARNING!
Update exclusively if the instrument has a fully charged battery!
FiberScanONE™
User
manual
rev.
1.0.3
Page
47 of 74