80
5. Accessing Calibration
Set
Calibration
in the capno waveform menu
to
Yes
using the jog dial. The message
“Calibration in progress” will appear.
6. Commencing Calibration Gas Flow
Once zero calibration has ended successfully, the message “Turn on CO
2
(10%)
calibration gas.” will appear. Commence calibration gas flow by turning the knob of the
flowmeter. The flow rate should be between 130~170 ml/min. The message “Calibration
gas detected” will appear.
7. Ending Calibration Gas
Once sampling is completed, the message “Please turn off CO
2
(10%) calibration gas” will
appear. Turn off the calibration gas by turning the knob of the flowmeter.
If Gas Calibration Fails
If gas calibration fails while in progress, a screen like the following message will be
displayed.
9
Zero calibration range error
9
Zero calibration signal unstable error.
9
High calibration range error.
9
High calibration signal unstable error.
In this case, gas calibration must be performed again, but before doing so, check the
following selections.
9
Is the correct calibration gas being used?
9
Are connections loose?
9
Is the calibration gas supply flow too low?
9
Is the calibration gas supply flow too high?
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...