To Initiate MANUAL Measurement Mode
1. Press the
NIBP Start/Stop button
.
A single blood pressure measurement will be made. The measurement will be displayed
for 180 minutes unless another measurement is initiated. A manual NIBP reading can be
obtained in AUTO mode by pressing
NIBP start/stop button
between two AUTO
measurements without the cancellation of AUTO mode.
To Initiate Automatic (AUTO) Measurement Mode
1. Press
NIBP interval button
to select the desired automatic mode interval from
NIBP
interval setting menu
(see Table 25). The initial measurement will start automatically
in a selected interval. The automatic mode can be also activated by pressing
NIBP
Start/Stop button
after selecting the interval.
9
NIBP Auto Mode intervals
: Off, Cont, 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60, 90, 120, 180 minutes
2. An NIBP reading can be cancelled by pressing
NIBP Start/Stop button
during the
AUTO measurements.
Note: When the time interval is set to 1 minute, the initial measurement will automatically
start after 5 seconds, and then the measurement interval will automatically become
2.5 minutes after 12 minutes elapsed.
The NIBP numerical area will display the
NIBP auto mode interval
and
NIBP elapsed
time icon
. The interval is the time from when one measurement starts to when the next
measurement starts. The measurement value will be displayed until another
measurement starts. When the AUTO mode is cancelled, the last measurement will be
displayed for 180 minutes.
In AUTO mode, the monitor attempts to meet the requirement of SVRP (Safe Venous
Return Pressure) as long as starting a new reading does not violate the requirement of
being 30 seconds below SVRP between readings. A new blood pressure reading will not
start until the 30 second period has elapsed. When CONT and 1 minute is selected in
NIBP interval setting menu
, this SVRP can be shorten over 2 seconds since CONT is
the intensive measurement during the short term which is 5 minutes in the BP-S510.
To Initiate Continuous (CONT) Measurement Mode
You may select
Cont
to activate the continuous measurement mode. The initial
measurement will automatically starts in 5 seconds. The measurement interval will
automatically become 2.5 minutes after 5 minutes elapsed. Also, if the
NIBP start/stop
button
is pressed during CONT mode, the measurement will be canceled and the interval
will be changed to 2.5 minutes.
To Stop Blood Pressure Measurements
You may press
NIBP Start/Stop button
at any time to stop the current measurement and
deflate the cuff. If an automatic measurement is underway, next measurement will start at
the next interval after the current measurement stops.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...