Alarm Messages
Check Items
SYSTEM: Failure.
Reboot the monitor. If the problem persists, cease use immediately
and contact qualified service personnel or your local supplier.
ECG : Check ECG leads &
electrodes.
ECG error is detected, Electrodes or lead wires may not be
correctly attached or a circuit is possibly saturated due to offset
voltage. Check whether electrodes are correctly attached and
electrodes are new and wet. Confirm the patient’s skin is clean.
{label}
: Cable/Sensor
disconnected.
Cable not connected to the main unit. If connected, the cable may
be damaged. Replace with a new cable. If replacing the cable has
no effect the problem may be within the device. In this case, cease
use immediately.
SpO
2
: Check probe.
Sensor is not in contact with patient. SpO
2
could not be measured.
Fit the sensor correctly to the patient, and measure again.
TEMP
{n}
: Temperature
probe disconnected.
Sensor not connected to the main unit. If connected, the cable may
be damaged. Replace with a new cable. If replacing the cable has
no effect the problem may be within the device. In this case, cease
use immediately.
CAPNO: Occlusion.
Check the patient condition.
CAPNO: Water trap full.
Disconnect the sampling tubing from the water trap and replace the
water trap with a new one.
ECG: Signal saturation.
Decrease the ECG size via the setup menu.
RESP: Check Resp leads
& electrodes.
Electrodes or lead wires may not be correctly attached or a circuit
is possibly saturated due to offset voltage.
Check whether electrodes are correctly attached and electrodes
are new and wet. Confirm the patient’s skin is clean.
{label}
: Unable to zero
calibration.
Could not zero calibrate pressure. Check that the transducer is
open to the atmosphere, and check the three-way tap. As it is also
possible that the measured pressure incorporates noise, check the
measurement circuit.
{label}
: Out of range.
A value outside the measurement range was obtained. As the
transducer has been subject to abnormal pressure, check the
measurement circuit.
SpO
2
: Check sensor.
Sensor not connected. If connected, the cable or connector may be
damaged. Replace with a new cable. If replacing the cable has no
effect the problem may be within the device. In this case, cease
use immediately.
SpO
2
: Sensor failure.
A problem with the SpO
2
sensor has been detected. The SpO
2
measurement function does not operate. The possible cause is a
connection failure of the SpO
2
sensor and the extension cable, or a
failure of the sensor or cable. Reconnect the sensor and extension
cable or replace them with new ones. If the problem doesn’t clear
up after carrying out the remedies above or switching power
OFF/ON, a grave fault can develop, Cease the use of the sensor
immediately.
SpO
2
: Module reset.
A problem with the SpO
2
measurement has been detected. The
SpO
2
measurement function does not operate. If switching power
OFF/ON has no effect it is possible that a fault has occurred.
Cease use immediately.
TEMP
{n}
: Out of range.
A measure reading outside the measurement range was obtained.
It is possible that the temperature in the vicinity of the sensor is
extremely low (less than 15.0°C) or extremely high (more than
45.0°C). Adjust the ambient temperature and measure again.
CAPNO: Zero calibration
range error
Try again the gas calibration.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...