46
Setup Connections
Note: Colin Medical Technology recommends the use of silver/silver chloride electrodes
(Ag/AgCl). When dissimilar metals are used for different electrodes, the electrodes
may be subject to large offset potentials due to polarization, which may be severe
enough to prevent obtaining an ECG trace. Using dissimilar metals may also
increase recovery time after defibrillation.
1. Select the electrodes to be used. Use only one type of electrode on the same patient to
avoid variations in electrical resistance. Prepare the electrode sites according to
electrode manufacturer’s instructions. See Figure 21 and 22 for electrode placement
configurations.
Figure 21. Standard 3 Electrode Placement
Figure 22. 5 Electrode Placement
Note: One of 5-1 to 5-6 Lead electrode placement sites for the fifth lead.
2. Connect the ECG lead wires and ECG cable.
3. Connect the ECG cable to the ECG connector on the monitor’s left panel.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...