110
Alarm Messages
Check Items
obtained. Check whether cuffs are not wrapped around thick
clothing. After wrapping cuffs around property, measure again.
When the error occurs in the initial measurement in continuous
mode, the second measurement will start unless Stop button is
pressed.
NIBP: Cuff excessive
artifact (C13)
Measurement failed because of patient movement during
measurement. Tell the patient to stay still, then, measure again.
When it occurs in the initial measurement in continuous mode, the
second measurement will start unless Stop button is pressed.
NIBP: Cuff insufficient
pressure (C14)
Measurement failed because of insufficient pressurizing. There is a
possibility that standard cuff pressure might be detected wrongly
due to noises, motion artifact or external vibration.
Check whether cuffs are not wrapped around thick clothing,
whether the patient is moving and whether cuffs are free from
outside vibrations, then, measure again.
When it occurs in the initial measurement in continuous mode, the
second measurement will start unless Stop button is pressed.
NIBP: Cuff irregular pulses
(C15)
Blood pressure could not be measured because oscillation graph
was not normal. There is a possibility that motion artifact or
vibration from outside might interrupt the measurement.
Check whether the patient stays still and cuffs are free from
external vibration, then, measure again.
When it occurs in the initial measurement in continuous mode, the
second measurement is continued unless Stop button is pressed.
NIBP: Cuff motion artifact
(C16)
Blood pressure could not be measured because noises interrupted
pulse waveform signal. Check for motion artifacts, or external
vibration and then, measure again.
When it occurs in the initial measurement in continuous mode, the
second measurement will start unless Stop button is pressed.
NIBP: Cuff time-out (C17)
Measurement was preventively stopped because measurement
time exceeded 160 seconds (adult), There is a possibility that
blood pressure might be repeatedly measured due to insufficient
pressurizing caused by calcified pseudohypertension.
NIBP: Cuff time-out, over
160 pulses (C18)
Pulse waveform signal more than 160 beats are detected during
measurement. There is possibility that noises might interrupt
signal.
Motion artifact or external vibrations possibly affected cuffs. Check
for patient movement and if the cuff is free from stays still and cuffs
are free from outside vibration, then, measure again.
NIBP: Cuff pressure failure
(C19)
Cuff pressure exceeded more than 300 mmHg (adult) during
measurement.
There is a possibility that the patient moved during measurement
or strong pressure from outside might be added to cuffs.
Considering above, measure again.
NIBP: Cuff weak pulse
(C20)
Amplitude of pulse obtained from cuffs are too weak.
This error possibly occurs when cuffs are wrapped around loosely
in ASO patients or when cuffs are wrapped around thick clothing.
Wrap cuffs around properly, then, measure again.
NIBP: Check cuff, hose
and mode (C21)
Patient to be measured, and cuff size used, do not match.
This error may occur if the blood pressure measurement mode
setting is incorrect, if the cuff has been wrapped tightly in the adult
mode, loosely in the neonatal mode or if the arm has been bent
during measurement.
Check the measurement mode setting and application of the cuff,
and measure again.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...