129
Manufacturer’s Declaration
WARNING: For best product performance and measurement accuracy, use only
accessories supplied or recommended by Colin Medical Technology. Use
accessories according to the manufacturer’s directions for use and your
facility’s standards. The use of accessories, transducers, and cables other than
those specified may result in increased emission and/or decreased immunity of
the BP-S510.
The BP-S510 is suitable for use in the specified electromagnetic environment. The
customer and/or user of the BP-S510 should assure that it is used in an
electromagnetic environment as described below;
Table 42. Electromagnetic Emissions (IEC60601-1-2)
Emission Test
Compliance
Electromagnetic Environment
RF emission
CISPR 11
Group 1
The BP-S510 must emit electromagnetic
energy in order to perform its intended
function. Nearby electronic equipment may
be affected.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emission
IEC 61000-3-3
Complies
The BP-S510 is suitable for use in all
establishments.
Table 43. Electromagnetic Immunity (IEC60601-1-2)
Immunity Test
IEC60601-1-2
Test Level
Compliance
Level
Electromagnetic
Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Electric fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply lines
±1 kV for
input/output lines
±2 kV for
power supply
lines
±1 kV for
input/output lines
Mains power quality should be
that of a typical commercial
and/or hospital environment
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV
common mode
±1 kV differential
mode
±2 kV
common mode
Mains power quality should be
that of a typical commercial
and/or hospital environment
Voltage dips,
short
interruptions and
voltage
variations on
power supply
<5 % U T
(>95 % dip in UT )
for 0.5 cycle
<5 % U T
(>95 % dip in U T)
for 0.5 cycle
IEC 61000-4-11
40 % U T
(60 % dip in UT )
for 5 cycles
40 % U T
(60 % dip in U T)
for 5 cycles
70 % U T
(30 % dip in UT )
for 25 cycles
70 % U T
(30 % dip in UT)
for 25 cycles
Mains power quality should be
that of a typical commercial
and/or hospital environment. If
the user of the BP-S510
requires continued operation
during power mains
interruption, it is recommended
that the BP-S510 be powered
from an uninterruptible power
supply or battery.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...