Informative messages indicate a system condition that needs to be corrected.
Table 15. Informative Messages
Parameter
Condition
Messages
Measurement value can’t be calculated
even when inflated cuff pressure is
deflated to specified pressure. Specified
pressure is 10mmHg for adult and 5
mmHg for neo. (C12)
NIBP: Retry, Check cuff/Patient (C12)
Deflation time is too long
Deflation speed is too low. (C13)
NIBP: Retry, Cuff excessive artifact (C13)
Maximum inflating pressure is too low to
calculate patient BP value. (C14)
NIBP: Retry, Cuff insufficient pressure
(C14)
Abnormal pulse due to arrhythmia and
too much noise. (C15)
NIBP: Retry, Cuff irregular pulses (C15)
Collected pulse’s movement is
abnormal. (C16)
NIBP: Retry, Cuff motion artifact (C16)
Number of detected pulse is beyond
specified number. Specified number is
160 for both adult and neo. (C18)
NIBP: Retry, Cuff time-out, over 160
pulses (C18)
Cuff pressure is beyond specified
pressure that is specified for patient
safety. Specified pressure is 300mmHg
for adult and 150mmHg for neo. (C19)
NIBP: Retry, Cuff pressure failure (C19)
NIBP
Cuff size is inadequate to patient. (C21)
NIBP: Retry, Check cuff, hose and mode
(C21)
ECG
Pacer pulse detection is on.
ECG: Pacer detect is on.
Waiting for zero calibration.
{label}
: No zero reading.
IBP
Zero calibration in progress.
{label}
: Zero CAL in progress.
Current measurement is affected by
patient motion.
SpO
2
: Motion artifact.
%SpO
2
On pulse searching
SpO
2
: Pulse search in progress.
Capno: Calibration in progress.
Capno: Turn on CO
2
(10%) calibration gas.
Capno: Turn off CO
2
(10%) calibration gas.
Messages about Zero calibration
process
Capno: Calibration gas detected.
Capno
Capno module is warming up.
Capno: Warming up in progress.
Recorder has no paper.
System: No recorder paper.
Device abnormally shut down last time
System: Abnormally shut down last time.
System
Device has no recorder module.
System: No recorder installed.
Alarm sound is silenced.
Audible alarm silenced.
Alarm sound is suspended.
Audible alarm suspended.
Alarm sound is inhibited.
Audible alarm inhibited.
Device in Demo Mode
Demo Mode.
Exit scrolling in trend screens
Press jog dial to exit scroll.
Other
Contact your service personnel.
Contact your service personnel.
Note: There may be other informative messages that are not listed above.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...