Respiration / Apnea Source
You can select either awRR or imRR for the source of the respiration rate. If you select
Auto, the monitor will automatically derive the respiration rate from one of the monitoring
parameters in this order of priority; awRR and imRR.
Note: You can select awRR as the source when CO
2
module is installed.
Apnea Time Setting
When the patient’s breath is not detected from the airway measurement for a selected
time setting, the monitor will activate an apnea alarm. When
Step 60
is selected, the
monitor will sound two beeps in 20 second interval from no breath. At 60 seconds elapsed,
an apnea alarm will be activated. When
Step 90
is selected, the monitor will sound three
beeps in 30 second interval from no breath. At 90 seconds elapsed, an apnea alarm will
be activated. If
Off
is selected, the monitor does not detect an apnea alarm.
When the monitor does not detect a respiration signal from the impedance measurement
for 40 seconds, the monitor will activate a loss of respiration alarm.
Check the condition of the patient, then check the connections of the patient cables.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...