![Colin BP-S510 Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/colin/bp-s510/bp-s510_operation-manual_2637159121.webp)
115
EMI (Electromagnetic Interference)
WARNING: Keep patients under close surveillance when monitoring. It is
possible, although unlikely, that radiated electromagnetic signals from sources
external to the patient and monitor can cause inaccurate measurement readings.
Do not rely entirely on the monitor readings for patient assessment.
WARNING: It is possible that any radio frequency transmitting equipment and
other nearby sources of electrical noise may result in disruption in the monitor
operation.
WARNING: It is possible, although unlikely, that large equipment using a
switching relay for its power on/off may affect monitor operation. Do not operate
the monitor in such environments.
This device has been tested and found to comply with the limits for medical devices to
the IEC60601-1-2, and the Medical Device Directive 93/42/EEC. These limits are
designed to provide reasonable protection against harmful interference in a typical
medical installation.
However, because of the proliferation of radio-frequency transmitting equipment and
other sources of electrical noise in health care environments (such as electrosurgical
equipment, defibrillator, cellular phones, mobile two-way radios, electrical appliances,
and high-definition television), it is possible that high levels of such interference due to
close proximity or strength of a source may affect monitor operation.
WARNING: The monitor is designed for use in environments in which the signal
can be obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the monitor may not seem to operate
correctly.
Monitor disruption may be indicated by erratic readings, cessation of operation, or other
incorrect functioning. If this occurs, survey the site to determine the source of this
disruption. Try the following actions to see if they eliminate the disruption:
z
Turn equipment in the vicinity off and on to isolate the offending equipment.
z
Reorient or relocate the interfering equipment.
z
Increase the separation between the interfering equipment and this equipment.
The monitor generates, uses, and can radiate radio frequency energy. If the monitor is
not installed and used in accordance with these instructions, the monitor may cause
harmful interference with other devices in the vicinity.
If assistance is required, contact your local supplier.
Obtaining Technical Assistance
For technical information and assistance, or to order a monitor service manual, call your
local supplier. The service manual provides information required by qualified service
personnel when servicing the monitor.
When calling your local supplier, you may be asked to provide the software version
number of your monitor. The software version is displayed when monitor power is
activated.
Содержание BP-S510
Страница 10: ...4 This page is intentionally left blank...
Страница 22: ...16 This page is intentionally left blank...
Страница 38: ...32 This page is intentionally left blank...
Страница 68: ...62 This page is intentionally left blank...
Страница 80: ...74 This page is intentionally left blank...
Страница 104: ...98 This page is intentionally left blank...
Страница 108: ...102 This page is intentionally left blank...
Страница 114: ...108 This page is intentionally left blank...
Страница 122: ...116 This page is intentionally left blank...
Страница 126: ...120 This page is intentionally left blank...