Installation
34
• make the electrical connection using the cable supplied with the generator to the
socket "M";
c
make sure that the off button on the rear of the generator is easily ac-
cessible
c
For reasons of safety, the following instructions must be strictly ob-
served.
The electrical installation must comply with current standards, in particular re-
garding the protection line.
c
Recommendations for correct installation:
• do not use extension cords, adapters or multiple sockets; if necessary, replace
the cable supplied with a longer one. If the supplied cable cannot be used, it is
the Customer's responsibility to obtain a suitable cable for the local configuration
of the electrical socket outlet;
• always connect the protection wire;
• the mains socket must be located in an easily accessible position;
• comply with the "IEC" power standards.
e
If a failure in gas supply (due to a power failure, activation of an electrical safety
device, or generator fault), even temporary, is not admissible, a pneumatic panel
should be envisaged, to enable provisional activation of a reserve gas source (to-
tally or partially automatic).
7.2.7.
Step 7: connect the vent hose
The drain outlet is a vent for the pure oxygen where condensation may also be
present.
• Locale the drain outlet "N" at the rear of the generator;
M
L