• in the screen shown above you have the option to choose whether to fill the in-
ternal tank and start production or fill the internal tank and stop the generator.
Until the internal tank is completely filled, approximately 3 minutes, the genera-
tor shows the following screen;
• if the quality of the water is not adequate, the following screens will automati-
cally be shown;
• the CPU prompts the procedure again from step 1;
• change the water in the reserve tank and perform the steps described above
again;
• when the demineralized water reaches the required values (see
par. 4 Demine-
ralized water)