Hydrogen generator
HyGen
55
tapping on the warning icon will display the following screen with Warning W03
active;
selecting the active warning will show a page explaining why it appeared;
now, clicking on "START SEQ NOW" will start the guided sequence for changing
the desiccant.
e
CAUTION! The procedure for replacing the desiccant, once begun, must be com-
pleted.
MATERIAL NECESSARY
TIME NECESSARY
Changing the desiccant takes 35 minutes of downtime, comprising:
15 minutes with the operator present to perform manual operations;
20 minutes of generator operation discharging hydrogen into the environment
to purify the dryer. This operation is designed not to pollute the line.
m
Always wear protective goggles, gloves and mask. Prevent the desic-
cant from coming into contact with moist skin and eyes
No.
Component
Code
1
set of 2 consumable refills for the Hydro and Hygen
series
422.93.0001