Hydrogen generator
HyGen
33
3.
remember to wash the internal tank thoroughly;
4.
after washing immediately fill the internal tank of the generator;
5.
on completing this process, keep the flask closed for future use;
• leave the tank open for the shortest possible time so as to decrease the possibi-
lity of polluting the water;
e
Do not use cloths or paper towels to clean the tank: they can release fibres and
particles that are harmful to the generator.
• screw on the cap of the internal tank;
• reposition the tank
e
Connect only jack "E", do not connect the quick connector "D"!
The connector "D" will be connected during the "first start-up" procedure
e
If the required type of water is not used, the performance of the generator may
no longer be ensured and it may lead to permanent damage to the generator
e
The maximum resistivity value displayed by the CPU is 1.5 m
Ω
cm @25°C
7.2.5.
Step 5: connect the hydrogen line
• locate the coupling "J" H2 DELIVERY at the rear of the generator;
• remove the cap "K";
e
Keep the cap along with the connection kit as it will be needed in the event of
uninstalling the generator
• connect the line.
e
The hydrogen line must be able to withstand a pressure of at least 10.2 bar
7.2.6.
Step 6: connect the electric mains
• Check that the mains specifications correspond to the data on the generator da-
taplate;
• ensure that the "POWER" button "L" is set to OFF;
J
K