VCC-2CXP6M
Rev.905-0186-02
©2019 CIS Corporation. All rights reserved.
14
Effective Line : 1248H
Effective Line
~
~
Exposure Out
1
2
4
8
1
TRIG IN
LVAL Out
FVAL Out
Video Out
~ ~
1
2
4
8
1
Effective Line
Trigger input ignore period
Exposure Time
Exposure Time Delay
(A)
TRIG IN
Exposure Out
Exposure time delay
Image
Header
Idle
Stream data
Stream data
Idle
Idle
Image
Header
Cxp Stream
5.9.
FAST Fixed Trigger Shutter Mode (ClockSync) CLK Sync. Trigger
(TriggerMode=On, TriggerSyncMode=ClockSync, TriggerActivation= RisingEdge)
□
This is the mode to start exposure with external input trigger signals, and execute exposure for set period.
□
Trigger operation is CLK sync., V-sync. reset system.
The delay time (Exposure Time Delay) from detecting the trigger edge in the camera to actually starting exposure is
approx. 0.13µs. Please input more than 1µs of pulse to the trigger.
□
There is an exposure time period for approx. 4.997µs at the edge right after exposure time.
SP OUT
(Exposure Out)
Actual exposure period
Approx. 4.997us
□
The next trigger while outputting video for the prior trigger cannot be accepted.
Trigger input while exposure period and reading out period shall be ignored in the camera. (Refer to (A) and (B) in the
drawing next page.)