Operation Overview
3-17
manually switch between the two programmed states. Slew rate determines the rate at
which Load level changes from one load level state to another. Figure 3-5 shows the current
level of load after pressing
A/ B
key.
State A=4A, State B=2A, Rise =0.2A/
μs, Fall =0.08A/μs
Time
Current
rise slew rate
fall slew rate
State A
State B
4A
2A
10uS
25uS
press key "A/B"
Figure 3-5 Load Level after Pressing
A/ B
Key
Constant Resistance Mode
3.6.2
In CR mode, the Load will sink a resistance in accordance with the programmed value
regardless of the input voltage. To enter into the CR mode, press the
MODE
key and select
CR
mode.
L 1 =
1 0 0 0 . 0 0
Ω
S R
=
0 . 0 0 5
A / u s
C C
L 2 =
1 0 0 0 . 0 0
Ω
S R
=
0 . 0 0 5
A / u s
▼
C R
0 . 0 0 0 V
0 . 0 0 0 A
C V
0 . 0 0 0 0
W
C P
C R _ H
V o n
=
0 . 0 0 0
V
L 1 =
1 0 0 0 . 0 0
Ω
I _ R A N G E =
H I H G ( 2 )
▲
C C
L 2 =
1 0 0 0 . 0 0
Ω
C R
0 . 0 0 0 V
0 . 0 0 0 A
C V
0 . 0 0 0 0
W
C P
C R _ H
V o n
=
0 . 0 0 0
V