High Power Electronic Load 63200A Series Operation & Programming Manual
3-32
To enter into EXTW mode, press
and select
EXTW
.
MODE: Able to set to CC, CR or CV mode.
CC mode
CR mode
CV mode
Ranges (Low, Middle, High)
It can be programmed in any of the three ranges, low range, middle range and high range.
The low range provides better resolution. If any value is over the maximum of low range,
you must select the middle range. When any value is over the maximum of middle range,
you must select the high range. To change the range, press the
RANGE
key few times until
the VFD range indicator is active at you want to select. The mode change will affect the Load,
so will the change of range. Both of them will cause the input to go through an off state.
The new setting will change the input immediately at a rate determined by the slew rate
setting.