Remote Operation
4-25
POWER Subsystem
4.3.2.7
POWer:STATic:L1
Type:
Channel-Specific
Description:
Set the static load power for constant power mode.
Setting Syntax:
POWer:STATic:L1<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: POW:STAT:L1 20
Set the load parameter L1 = 20W.
POW:STAT:L1 10W
Set the load parameter L1 = 10W.
POW:STAT:L1 MAX
Set the load parameter L1 = maximum
value.
POW:STAT:L1 MIN
Set the load parameter L1 = minimum
value.
Query Syntax:
POWer:STATic:L1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Watt]
Query Example:
POW:STAT:L1?
POW:STAT:L1? MAX
POW:STAT:L1? MIN
POWer:STATic:L2
Type:
Channel-Specific
Description:
Set the static load power for constant power mode.
Setting Syntax:
POWer:STATic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: POW:STAT:L2 20
Set the load parameter L2 = 20W.
POW:STAT:L2 10W
Set the load parameter L2 = 10W.
POW:STAT:L2 MAX
Set the load parameter L2 = maximum
value.
POW:STAT:L2 MIN
Set the load parameter L2 = minimum
value.
Query Syntax:
POWer:STATic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Watt]
Query Example:
POW:STAT:L2?
POW:STAT:L2? MAX
POW:STAT:L2? MIN
POWer:STATic:RISE
Type:
Channel-Specific
Description:
Set the rising slew rate of current for constant power mode.
Setting Syntax:
POWer:STATic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: POW:STAT:RISE 2.5
Set rising slew rate to 2.5A/μs.
POW:STAT:RISE 1A/μs
Set rising slew rate to 1A/μs .
POW:STAT:RISE MAX
Set rising slew rate to the
maximum value of load.
POW:STAT:RISE MIN
Set rising slew rate to the
minimum value of load.
Query Syntax:
POWer:STATic:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>,
[Unit = A/μs]
Query Example:
POW:STAT:RISE?
POW:STAT:RISE? MAX
POW:STAT:RISE? MIN
Содержание 63200A Series
Страница 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Страница 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Страница 7: ...vi ...
Страница 11: ......
Страница 15: ......
Страница 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Страница 23: ......
Страница 31: ......
Страница 127: ......