High Power Electronic Load 63200A Series Operation & Programming Manual
4-22
RESistance:DYNamic:T2
Type:
Channel-Specific
Description:
Set duration parameter T2 for constant resistance dynamic mode.
Setting Syntax:
RESistance:DYNamic:T2<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 10μs ~ 100s, Resolution = 10μs, Unit = Second
Setting Example: RES:DYN:T2 10ms
Set the dynamic duration T2 = 10ms.
RES:DYN:T2 2
Set the dynamic duration T2 = 2s.
RES:DYN:T2 MAX
Set the dynamic duration T2 as
maximum value.
RES:DYN:T2 MIN
Set the dynamic duration T2 as
minimum value.
Query Syntax:
RESistance:DYNamic:T2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example:
RES:DYN:T2?
RES:DYN:T2? MAX
RES:DYN:T2? MIN
RESistance:DYNamic:REPeat
Type:
Channel-Specific
Description:
Set the repeat count for constant resistance dynamic mode.
Setting Syntax:
RESistance:DYNamic:REPeat<space><NRf+>
Setting Parameters:<NRf+>, 0 ~ 65535, Resolution = 1, Unit = None
Setting Example: RES:DYN:REP 500
Set repeat count = 500
RES:DYN:REP MAX
Set repeat count = maximum value.
RES:DYN:REP MIN
Set repeat count = minimum value.
Query Syntax:
RESistance:DYNamic:REPeat?[<space><MAX | MIN>]
Return Parameters:<NR1>, [Unit = None]
Query Example:
RES:DYN:REP?
RES:DYN:REP? MAX
RES:DYN:REP? MIN
RESistance:DYNamic:RISE
Type:
Channel-Specific
Description:
Set the rising slew rate of current for constant resistance dynamic
mode.
Setting Syntax:
RESistance:DYNamic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:DYN:RISE 2.5
Set rising slew rate to 2.5A/μs.
RES:DYN:RISE 1A/μs
Set rising slew rate to 1A/μs.
RES:DYN:RISE MAX
Set rising slew rate to the maximum
value of dynamic load.
RES:DYN:RISE MIN
Set rising slew rate to the minimum
value of dynamic load.
Query Syntax:
RESistance:DYNamic:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>,
[Unit = A/μs]
Query Example:
RES:DYN:RISE?
RES:DYN:RISE? MAX
RES:DYN:RISE? MIN
RESistance:DYNamic:FALL
Type:
Channel-Specific
Description:
Set the falling slew rate of current for constant resistance dynamic
mode.
Содержание 63200A Series
Страница 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Страница 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Страница 7: ...vi ...
Страница 11: ......
Страница 15: ......
Страница 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Страница 23: ......
Страница 31: ......
Страница 127: ......