Remote Operation
5-5
Standard Event Status
5.2.5
■
All programming errors that have occurred will set one or more error bits in the Standard
Event Status register. Table 5-3 describes the standard events that apply to the
electronic load.
■
Reading the Standard Event Status register will reset it to zero.
■
The Standard Event Enable register can be programmed to specify the standard event
bit that is logically ORed to become Bit 5 (ESB bit) in the Status Byte register.
Table 5-3 Bit Description of Standard Event Status
Mnemonic Bit Value
Meaning
OPC
0
1
Operation Complete.
This event bit generated is responding to
the *OPC command. It indicates that the device has completed
all of the selected pending operations.
QYE
2
4
Query Error
. The output queue was read when no data were
present or the data in the queue were lost.
DDE
3
8
Device Dependent Error
.
Memory was lost, or self-test failed.
EXE
4
16
Execution Error
. A command parameter was out of the legal
range or inconsistent with the electronic load’s operation, or the
command could not be executed due to some operating
conditions.
CME
5
32
Command Error
. A syntax or semantic error has occurred, or the
electronic load has received a <GET> message from program.
Status Byte Register
5.2.6
■
The Status Byte register summarizes all of the status events for all status registers.
Table 5-4 describes the status events that are applied to the electronic load.
■
The Status Byte register can be read with a serial of pull or *STB? query.
■
The RQS bit is the only bit that is automatically cleared after a serial of pull.
■
When the Status Byte register is read with a *STB? query, Bit 6 of the Status Byte
register will contain the MSS bit. The MSS bit indicates that the load has at least one
reason for requesting service. *STB? does not affect the status byte.
■
The Status Byte register is cleared by *CLS command.
Table 5-4 Bit Description of Status Byte
Mnemonic Bit Value
Meaning
CSUM
2
4
Channel Summary
. It indicates if an enabled channel event has
occurred. It is affected by Channel Condition, Channel Event and
Channel Summary Event registers.
QUES
3
8
Questionable
. It indicates if an enabled questionable event has
occurred.
MAV
4
16
Message Available.
It indicates if the Output Queue contains
data.
ESB
5
32
Event Status Bit
. It indicates if an enabled standard event has
occurred.
RQS/MSS
6
64
Request Service/Master Summary Status
. During a serial of pull,
RQS is returned and cleared. For a *STB? query, MSS is
returned without being cleared.
Содержание 63200A Series
Страница 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Страница 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Страница 7: ...vi ...
Страница 11: ......
Страница 15: ......
Страница 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Страница 23: ......
Страница 31: ......
Страница 127: ......