High Power Electronic Load 63200A Series Operation & Programming Manual
4-36
Setting Example: CURR:SWE:DWEL 50
Set dwell time = 50s
CURR:SWE:DWEL 500ms
Set dwell time = 0.5s
CURR:SWE:DWEL MAX
Set dwell time = maximum value.
CURR:SWE:DWEL MIN
Set dwell time = minimum value.
Query Syntax:
[ADVance:]CURRent:SWEep:DWELl?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example:
CURR:SWE:DWEL?
CURR:SWE:DWEL? MAX
CURR:SWE:DWEL? MIN
[ADVance:]CURRent:SWEep:DUTY
Type:
Channel-Specific
Description:
Set the duty cycle for constant current frequency sweep mode.
Setting Syntax:
[ADVance:]CURRent:SWEep:DUTY<space><NRf+>
Setting Parameters:<NRf+>, 1% ~ 99%, Resolution = 1%
Setting Example: CURR:SWE:DUTY 50
Set duty cycle = 50%
CURR:SWE:DUTY MAX
Set duty cycle = maximum value.
CURR:SWE:DUTY MIN
Set duty cycle = minimum value.
Query Syntax:
[ADVance:]CURRent:SWEep:DUTY?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = None]
Query Example:
CURR:SWE:DUTY?
CURR:SWE:DUTY? MAX
CURR:SWE:DUTY? MIN
[ADVance:]CURRent:SWEep:RISE
Type:
Channel-Specific
Description:
Set the rising slew rate of current for constant current frequency
sweep mode.
Setting Syntax:
[ADVance:]CURRent:SWEep:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:SWE:RISE 2.5
Set rising slew rate to 2.5A/μs.
CURR:SWE:RISE 1A/μs
Set rising slew rate to 1A/μs.
CURR:SWE:RISE MAX
Set rising slew rate to the maximum
value of static load.
CURR:SWE:RISE MIN
Set rising slew rate to the minimum
value of static load.
Query Syntax:
[ADVance:]CURRent:SWEep:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>,
[Unit = A/μs]
Query Example:
CURR:SWE:RISE?
CURR:SWE:RISE? MAX
CURR:SWE:RISE? MIN
[ADVance:]CURRent:SWEep:FALL
Type:
Channel-Specific
Description:
Set the falling slew rate of current for constant current frequency
sweep mode.
Setting Syntax:
[ADVance:]CURRent:SWEep:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:SWE:FALL 2.5
Set falling slew rate to 2.5A/μs.
CURR:SWE:FALL 1A/μs
Set falling slew rate to 1A/μs.
CURR:SWE:FALL MAX
Set falling slew rate to the
maximum value of static load.
CURR:SWE:FALL MIN
Set falling slew rate to the
Содержание 63200A Series
Страница 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Страница 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Страница 7: ...vi ...
Страница 11: ......
Страница 15: ......
Страница 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Страница 23: ......
Страница 31: ......
Страница 127: ......